Model Number 101-9812 |
Device Problems
Difficult to Open or Close (2921); Material Integrity Problem (2978)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/28/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that during a superion indirect decompression system implant procedure while the physician was attempting to open the device in an extremely tight space the bracket around the screw came off of the implant.The physician noted an abnormal amount of osteophytes which made it challenging to open the device.The broken implant was replaced and the procedure was completed successfully.
|
|
Manufacturer Narrative
|
The returned spacer was analyzed and revealed that the spindle cap was completely sheared off from the implant body.This damage to the spacer indicates failure was likely due to deployment against resistance, such as, bone, and or manipulation of the position of the device by gear shifting of the inserter.
|
|
Event Description
|
It was reported that during a superion indirect decompression system implant procedure while the physician was attempting to open the device in an extremely tight space the bracket around the screw came off of the implant.The physician noted an abnormal amount of osteophytes which made it challenging to open the device.The broken implant was replaced and the procedure was completed successfully.
|
|