This mdr was generated under protocol b10009704 as a result of warning letter cms# (b)(4).A device history review was performed subsequent to the manufacturing of the device and prior to its release.No problems or issues were identified during this device history review.Visual inspection observed that the enclosure was dirty and scuffed.Line cord appears newer.Power switch is jammed into the device.Contains old style printed circuit board and drain fitting.During electrical testing and visual inspection, the reported problem was duplicated.The root cause of the reported issue was found to be the grounding nuts were loose inside the device.No action taken due to the age and condition of the device.D4 (udi) is unknown.No product information has been provided to date.
|