OAKDALE LEVEL 1 HOTLINE FLUID WARMER; WARMER, THERMAL, INFUSION FLUID Back to Search Results
Catalog Number CON-HL-90
Device Problem Electrical /Electronic Property Problem (1198)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
It was reported that there was a ground fault.No patient injury was reported.
 
Manufacturer Narrative
This mdr was generated under protocol b10009704 as a result of warning letter cms# (b)(4).A device history review was performed subsequent to the manufacturing of the device and prior to its release.No problems or issues were identified during this device history review.Visual inspection observed that the enclosure was dirty and scuffed.Line cord appears newer.Power switch is jammed into the device.Contains old style printed circuit board and drain fitting.During electrical testing and visual inspection, the reported problem was duplicated.The root cause of the reported issue was found to be the grounding nuts were loose inside the device.No action taken due to the age and condition of the device.D4 (udi) is unknown.No product information has been provided to date.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 HOTLINE FLUID WARMER
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
OAKDALE
3350 granada ave n
oakdale MN 55128
Manufacturer (Section G)
OAKDALE
3350 granada ave n
oakdale MN 55128
Manufacturer Contact
jim vegel
6000 nathan lane north
on condell dr
minneapolis, MN 55442
MDR Report Key15650933
MDR Text Key302205816
Report Number3012307300-2022-25712
Device Sequence Number1
Product Code LGZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K001764
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial
Report Date 10/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/21/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberCON-HL-90
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/31/2020
Was the Report Sent to FDA? No
Date Manufacturer Received09/28/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/09/2006
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-