|
Catalog Number 05.001.201 |
Device Problem
Break (1069)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/01/2022 |
Event Type
malfunction
|
Event Description
|
It was reported from (b)(6) that during service and evaluation, it was determined that the battery handpiece device failed leak tightness test, had component damage, the trigger moving parts did not move smoothly, would not hold/secure battery, the moving parts did not move smoothly and would not run.It was further determined that the device failed pretests for leakage test using bubble emission technique, mechanical free moving, sticky triggers, falling out protection and function of device.It was noted in the service order that the device switch was defective.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
|
Manufacturer Narrative
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.The device was returned for evaluation.During repair, an evaluation was performed and it was determined that the reported condition was confirmed.The assignable root cause was traced to component failure due to normal wear.Udi: (b)(4).
|
|
Search Alerts/Recalls
|
|
|