Model Number SL-2010M2096 |
Device Problem
Leak/Splash (1354)
|
Patient Problem
Hemorrhage/Bleeding (1888)
|
Event Date 10/06/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
This report has been identified as b.Braun medical internal report number (b)(4).The investigation is ongoing at this time.A follow up will be submitted when the investigation results become available.
|
|
Event Description
|
As reported by the user facility: event 2.Detailed inquiry description: arterial dialyzer connector snapped off of bloodline, one during treatment and one during setup.Lost 200 ml of blood, unable to return blood to patient.Treatment interrupted due to break in bloodline and to do new setup." no injury reported.
|
|
Manufacturer Narrative
|
Event 2: this report has been identified as b.Braun medical internal report number (b)(4).No sample was provided for evaluation.Based on the data from the investigation results, the root cause of the reported incident was unable to be determine.The actual defective device is a valuable tool in investigating the cause of this incident.A review of manufacturing records was performed and indicated that there were no quality issues during the manufacturing of this lot related to the reported issue.We will maintain this report for further references and continue to monitor other reports for similar occurrences.If any additional pertinent information becomes available, a follow up will be submitted.
|
|