|
Model Number SL-2010M2096 |
Device Problem
Air/Gas in Device (4062)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/26/2022 |
Event Type
malfunction
|
Event Description
|
As reported by the user facility: event 10: detailed inquiry description - kinked, twisted tubing, at times causing air bubbles.No injury reported.
|
|
Manufacturer Narrative
|
This report has been identified as b.Braun medical internal report number (b)(4).The investigation is ongoing at this time.A follow up will be submitted when the investigation results become available.
|
|
Manufacturer Narrative
|
Event 10: this report has been identified as b.Braun medical internal report number (b)(4).One sample was evaluated.Based on the evaluation results, no defects were observed.The reported defect was not confirmed.A review of manufacturing records was performed and indicated that there were no quality issues during the manufacturing of this lot related to the reported issue.We will maintain this report for further references and continue to monitor other reports for similar occurrences.If any additional pertinent information becomes available, a follow up will be submitted.
|
|
Search Alerts/Recalls
|
|
|