• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. LIGAMAX-5MM ENDO CLIP APPLIER; CLIP, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. LIGAMAX-5MM ENDO CLIP APPLIER; CLIP, IMPLANTABLE Back to Search Results
Model Number EL5ML
Device Problem Mechanical Jam (2983)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/05/2022
Event Type  malfunction  
Event Description
It was reported that during a laparoscopic cholecystectomy, clip was not fed into the jaws.Another device was used to complete the case.There were no adverse consequences to the patient.
 
Manufacturer Narrative
(b)(4).Batch # x9494c.Investigation summary: the product was returned to ethicon for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that the el5ml device was returned with no damage in the external components; upon visual inspection, four clips were noted to be jammed.The clips were removed in order to perform functional testing.In addition, the tyvek was returned along with the instrument.In an attempt to replicate the reported incident, the device was tested for functionality.Upon testing, it was noted that the clips were not fed into the jaws in the next actuation of the trigger.In order to evaluate the condition of the internal components of the device, it was disassembled.Upon disassembling, the feed link was noted broken causing the feeding issues and 9 clips were found inside the clip track.The event reported was confirmed and it is related to improper use of the device.Possible causes for the damage found is due to the jaws may have been restricted during firing, or not fully through the trocar during firing.Please reference the instruction for use for more information.As part of ethicon¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device lot and batch number, and no non-conformances were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIGAMAX-5MM ENDO CLIP APPLIER
Type of Device
CLIP, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*  00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*   00969
Manufacturer Contact
kate karberg
475 calle c
guaynabo 
*  
3035526892
MDR Report Key15653117
MDR Text Key307008298
Report Number3005075853-2022-07123
Device Sequence Number1
Product Code FZP
UDI-Device Identifier10705036001843
UDI-Public10705036001843
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K050344
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Other
Type of Report Initial
Report Date 10/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberEL5ML
Device Catalogue NumberEL5ML
Device Lot NumberX9494C
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/29/2022
Initial Date Manufacturer Received 10/03/2022
Initial Date FDA Received10/21/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/09/2022
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-