Investigation result: the subject device was not returned to olympus for the following reasons: the customer discarded the subject device.The subject device was not returned to olympus.Therefore, the reported phenomenon and condition of the device could not be confirmed.The subject device was manufactured on november 2019 based on the provided 3 digit lot information 9yk¿.A review of the device history record found the subject device was shipped in accordance with specifications.The record review of the dhr includes the following.Process inspection sheet.Quality inspection sheet.Nonconforming product report.Instruction for use (ifu) : the instruction manual contains the following descriptions, and it warns against this event.(rk1556 rev.01).The operator of this instrument must be a physician or medical personnel under the supervision of a physician and must have received sufficient training in clinical endoscopic technique.This manual does not explain or discuss clinical endoscopic procedures.Straighten out the instrument before inspecting it.The instrument can be damaged if it is coiled while the handle is operated.Operate the slider slowly, otherwise the tube could buckle.When inserting the instrument into the endoscope, retract the needle into the sheath, hold the instrument close to the biopsy valve, and keep it as straight as possible relative to the biopsy valve.Otherwise, the instrument could be damaged.Insert the instrument slowly.Abrupt insertion could damage the endoscope and/or instrument.Stop using the instrument if the insertion portion bends excessively during use.This could result in malfunction, such as failing to extend the needle or inject a fluid.Conclusion summary: the exact cause of the reported issue could not be conclusively identified by the following reason; the device was not retuned for the investigation.No abnormalities on the device was considered as no abnormalities were found in the device history record.From a similar complaint in the past, it was likely that the phenomenon ¿unable to inject liquid into the target tissue¿ occurred due to the compressive bucking on the needle tube.The compressive buckling on the needle tube was likely caused when the needle was extended due to the great friction between the outer tube and the needle.It was likely that the friction between the outer tube and the needle increased by the following factors.The needle extended/retracted while the tube was coiled in inspection of operation.The slider was abruptly pushed.The kink of the tube.Angle of the distal end of the endoscope.Olympus will continue to monitor complaints for this device.
|
It was reported during a therapeutic endoscopy procedure, the nurse opened and inserted the needle nm-401l-0423 into the instrument channel and when tried to inject liquid, it could not be injected as "it was covered " per the reporter.The same device was not used to complete the procedure.The intended procedure was completed using another device.There was no patient harm, no user injury reported due to the event.
|