• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAILLEFER INSTRUMENTS HOLDING SARL X-SMART IQ BASIC STARTER KIT; CONTROLLER, FOOT, HANDPIECE AND CORD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAILLEFER INSTRUMENTS HOLDING SARL X-SMART IQ BASIC STARTER KIT; CONTROLLER, FOOT, HANDPIECE AND CORD Back to Search Results
Catalog Number A105400000000
Device Problems Detachment of Device or Device Component (2907); Mechanics Altered (2984)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
There has been a previous report received with a similar device where this malfunction resulted in a serious injury.Therefore, it must be presumed that recurrence of this malfunction could possibly cause or contribute to a serious injury or require medical or surgical intervention to preclude such.As such, this event is reportable per 21cfr part 803.The device is available for evaluation, though results are not available as of this report.Evaluation results will be submitted as they become available.
 
Event Description
In this event it is reported that a x-smart iq basic starter kit won't hold files.No injury occurred.
 
Manufacturer Narrative
Investigation results: received: 1x x-smart iq contra-angle a105100000000 sn: (b)(6).X-smart iq ca head housing sav bad maintenance (user) rusty.Improper / inadequate cleaning, maintenance replaced 1x iq contra-angle a105100000000 40940.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
X-SMART IQ BASIC STARTER KIT
Type of Device
CONTROLLER, FOOT, HANDPIECE AND CORD
Manufacturer (Section D)
MAILLEFER INSTRUMENTS HOLDING SARL
ch. du verger 3
ballaigues vaud CH-13 38
SZ  CH-1338
Manufacturer (Section G)
MAILLEFER INSTRUMENTS HOLDING SARL
ch. du verger 3
ballaigues vaud CH-13 38
SZ   CH-1338
Manufacturer Contact
hannah seevaratnam
221 west philadelphia st.
york, PA 17401
7178457511
MDR Report Key15657383
MDR Text Key302801981
Report Number8031010-2022-00665
Device Sequence Number1
Product Code EBW
Combination Product (y/n)N
Reporter Country CodeSZ
PMA/PMN Number
K161213
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date 11/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberA105400000000
Device Lot Number35554
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/17/2022
Is the Reporter a Health Professional? Yes
Distributor Facility Aware Date10/17/2022
Initial Date Manufacturer Received 10/17/2022
Initial Date FDA Received10/24/2022
Supplement Dates Manufacturer Received10/17/2022
Supplement Dates FDA Received11/30/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Type of Device Usage A
Patient Sequence Number1
-
-