The lower edge of the pedicle on both sides of l5 fractured during screw insertion in the cortical bone trajectory.On the left, the lower edge of the pedicle was slightly cracked, and on the right, the entire lower edge of the pedicle was cracked.Bone quality was normal.The surgeon used myspinemc guide.He did a 5mm under drill, then a 5mm tap with a guide wire, and then a 6 mm tap.One side still seemed to be fixed, so the screw was left as it was.The fracture on the other side was large and the larger screw was inserted into the pedicle screw trajectory.
|
Clinical evaluation performed by medical affairs director: this complaint refers to a lumbar stabilization surgery.The surgeon planned for a 6 mm screw (comparatively large) to be placed very near the edge of the pedicles: he was most probably seeking to give the best possible stability, using a cortical-based trajectory and the largest possible screw.The planning system warned that the screws were likely to breach the pedicle but, certainly for medical reasons, the surgeon decided to maintain the plan.The patient-specific guides were manufactured accordingly.In these conditions, one of the pedicles fractured but the screw was judged to be soundly placed anyway.On the other side, the fracture of the pedicle limited the stability of the screw and therefore the surgeon decided to follow the traditional divergent path.No adverse clinical consequence should be expected from this change of surgical strategy, justified by the conditions detected during surgery.No reason to suspect a malfunctioning device.Visual inspection performed by r&d spine manager: during visual inspection, the pedicle screw (ref.(b)(4), lot 2126415) has been analyzed and is conform to specification.The myspine planning has been reviewed and no anomalies were found.The reported event may be related to many factors, including the need to place the screw very close to the cortical, increasing the likelihood of bone fracture.No definitive root cause can be established.
|