• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEXCOM, INC. DEXCOM G6 CONTINUOUS GLUCOSE MONITORING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEXCOM, INC. DEXCOM G6 CONTINUOUS GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number MT24078
Device Problem Unexpected Shutdown (4019)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/06/2022
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that unexpected receiver shutdown occurred.Product has been received but is pending evaluation.A follow up report will be submitted upon completion.No injury or medical intervention was reported.
 
Manufacturer Narrative
(b)(4).
 
Event Description
The product was evaluated.An external visual inspection was performed and failed due to usb connector damage.Pictures were taken.Receiver charge and boot test were performed and failed due to the receiver not turning on.Functional tests were not performed due to the receiver not turning on.An internal visual inspection was performed and failed due to moisture damage.Pictures were taken.The receiver log was not downloaded for reviewdue to the receiver not turning on.The allegation was undetermined.The probable cause could not be determined.No injury or medical intervention was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEXCOM G6 CONTINUOUS GLUCOSE MONITORING SYSTEM
Type of Device
CONTINUOUS GLUCOSE MONITOR
Manufacturer (Section D)
DEXCOM, INC.
6340 sequence dr.
san diego CA 92121
Manufacturer Contact
ashley spoto
6340 sequence dr.
san diego, CA 92121
8582000200
MDR Report Key15667394
MDR Text Key306718149
Report Number3004753838-2022-198746
Device Sequence Number1
Product Code QDK
UDI-Device Identifier00386270000583
UDI-Public00386270000583
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
DEN170088
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMT24078
Device Catalogue NumberSTR-OR-001
Device Lot NumberP12732080
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/19/2022
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 10/06/2022
Initial Date FDA Received10/25/2022
Supplement Dates Manufacturer Received11/03/2022
Supplement Dates FDA Received11/18/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/18/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age53 YR
Patient SexFemale
Patient Weight77 KG
-
-