This event has been recorded by zimmer biomet under (b)(4).This medwatch is being filed to relay additional information.The following sections have been updated: b4, g3, g6, h2, h3, h4, h6, h8, h10.Review of the most recent repair record determined the unit was found to not power on, the calibration was out, the switch was loose and the screw was missing.The motor, switch pack, ball bearings, and needle bearings were replaced and resolved the reported issue.Dhr was reviewed and no discrepancies related to the reported event were found.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|