Model Number 55740106550 |
Device Problem
Break (1069)
|
Patient Problem
Device Embedded In Tissue or Plaque (3165)
|
Event Date 09/24/2022 |
Event Type
Injury
|
Manufacturer Narrative
|
Implant date is unknown.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Information was received from a healthcare professional via manufacturer representative regarding a screw used in navigated fusion.Levels implanted: l4 - s1.It was reported that the patient fell from the toilet onto the bathroom floor and the screw broke midway at shaft.Part of it is still in the bone.A revision l4-s1 was performed.Patient medical history includes asthma, obesity, chronic kidney disease, hypertension, gerd, auto immune disease, immunosuppressive, degenerative joint disease (lumbar), leg weakness, and history of systemic lupus erythematosus.Patient symptoms are unknown.No further complications were reported/anticipated.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
H3.Product analysis #705259650: product# 55740106550 lot# h5748632 visual examination of the mas bone screw confirmed bone screw complete fracture at approximately ~3 threads from the tip of the bone screw.Optical examination did not identify material defect near the area of fracture origin, which could contribute to crack propagation.Microscopic examination of the fracture surface identified gently curving convex striations through the cross-sectional area of the implant, which is consistent with overload until subsequent mechanical failure of the implant.Additional information radiographic image review: 1.Ap x-ray posterior lumbar fusion 2.Lateral x-ray l4-s1 fusion a fractured screw tip is present on one side of l4, unclear if new hardware has been placed in provided image.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|