Model Number U3513084
Device Problems Inflation Problem (1310); Material Rupture (1546)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/06/2022
Event Type  malfunction  
Event Description
It was reported that during an angioplasty procedure via femoral artery, the pta balloon allegedly had inflation issue.The procedure was completed using another balloon.There was no reported patient injury.
 
Manufacturer Narrative
As the lot number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.Expiry date: 03/2024.
 
Manufacturer Narrative
H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: one ultraverse 035 pta dilatation catheter was returned for evaluation.No specific anomalies were noted during the visual evaluation.During the functional testing, the balloon was successfully inflated using the in-house presto inflation device.However, under microscopic observation, a pinhole rupture was noted on the balloon.Although during the functional testing the balloon inflated successfully, water was leaking from the balloon; therefore, investigation for the reported inflation issue was confirmed as water was also noted to be leaking from the pin-hole balloon during inflation.The investigation was also confirmed for the identified balloon rupture, as during the inflation of the device, a pinhole rupture was noted, which was also confirmed under microscopic observation.A definitive root cause for the reported inflation issue and identified pin-hole balloon rupture could not be determined based upon the provided information.Labeling review: a review of product labeling documents (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, and unit label) showed that the product labeling is adequate.H10: d4 (expiry date: 03/2024).H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event Description
It was reported that during an angioplasty procedure via femoral artery, the pta balloon allegedly had inflation issue.The procedure was completed using another balloon.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRAVERSE 035
Type of Device
PTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
Manufacturer (Section G)
CLEARSTREAM TECHNOLOGIES LTD.
moyne upper
enniscorthy, co. wexford N A
EI   N A
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key15672749
MDR Text Key302422580
Report Number2020394-2022-00839
Device Sequence Number1
Product Code LIT
UDI-Device Identifier00801741094170
UDI-Public(01)00801741094170
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K142261
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberU3513084
Device Catalogue NumberU3513084
Device Lot NumberCMFR0348
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/23/2022
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/29/2022
Initial Date FDA Received10/26/2022
Supplement Dates Manufacturer Received11/24/2022
Supplement Dates FDA Received12/05/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/22/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-