• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) GALLANT HF; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) GALLANT HF; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number CDHFA500Q
Device Problems Premature Discharge of Battery (1057); Interrogation Problem (4017)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/06/2022
Event Type  malfunction  
Event Description
It was reported that the patient presented for an implant procedure.During the procedure, the implantable cardioverter defibrillator to be implanted exhibited premature battery depletion and interrogation problem.The device was not implanted.
 
Manufacturer Narrative
The reported event of premature battery depletion was confirmed.Interrogation of the device revealed the device was above elective replacement indicator (eri) when received.Analysis of device image found the device was drawing high current.Electrical testing confirmed high current drain from the hybrid.An anomalous integrated circuit (ic) inside the hybrid was found to be the cause of the high current drain and premature battery depletion.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GALLANT HF
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key15674247
MDR Text Key305626793
Report Number2017865-2022-43138
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05415067032010
UDI-Public05415067032010
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/19/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/31/2024
Device Model NumberCDHFA500Q
Device Catalogue NumberCDHFA500Q
Device Lot NumberS000085245
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/06/2022
Initial Date FDA Received10/26/2022
Supplement Dates Manufacturer Received01/31/2023
Supplement Dates FDA Received02/20/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/20/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-