The datalogs were reviewed and based on the evaluation of the data, the handpiece and system performed as expected.The system has software safeguards (such as a power on self-test) that will trigger error/event codes should the system be outside of acceptable limits.The tip was evaluated, and the tip failed the leak test.The tip also failed visual inspection for disturbed glue in one corner.No dielectric breakdown was observed.The datalogs also showed that the tip had an issue with the cryogen distribution.The thermistor 1 appeared warmer than the other thermistors; however, the tip passed the thermistor test.No functional testing was performed due to the tip being expired.It was reported that the patient was given iv superficial anesthetic for the treatment.The patient should never be placed under anesthetics during thermage flx treatment.The patient must be alert and provide required feedback during treatment.Patient feedback regarding their perception of heat or discomfort during the procedure is an essential input to guide the operator in determining safe and effective treatment levels.Based on the available information, this treatment was performed in an off-label manner that resulted in unsafe conditions for the patient.A review of the manufacturing records showed all requirements were met.The complaint type is identified within the risk analysis.The lot history, trend analysis, risk analysis and/or directions for use review were considered acceptable, with the product performing within anticipated rates.Final test verification specifications are acceptable.No non-conformities or anomalies found related to this complaint when reviewing the device history record.No corrective action is required.
|