SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX ANESTHESIA BREATHING CIRCUIT; CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE)
|
Back to Search Results |
|
Catalog Number C37101307J |
Device Problem
Gas/Air Leak (2946)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/01/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
Other, other text: a product sample was received and is awaiting evaluation and investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.Udi number and g5 are unknown, no product information has been provided to date.
|
|
Event Description
|
It was reported that leakage of air from the product was observed, during the pre-use check.No patient injury was reported.
|
|
Manufacturer Narrative
|
A device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.A product sample and one (1) picture was received for evaluation.Visual inspection found tier damage due embedded resin in the circuit; thus, failure mode is confirmed.Other analysis was not required.Per trend review as leakage failure an escalation was performed to investigate this failure and determine the root cause through an internal capa.The principal cause defined by the team is the screw and barrel wear on flights, condition that affect directly to the internal plastination process due worn flights on screw.Current capa status ? implementation results.This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).
|
|