Model Number EL5ML
Device Problems Failure to Form Staple (2579); Failure to Fire (2610); Mechanical Jam (2983)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/01/2022
Event Type  malfunction  
Event Description
It was reported that during an unknown procedure, the staple was stuck and broke.
 
Manufacturer Narrative
(b)(4).Date of event unknown, assumed first day of month that complaint was reported batch # x9506v.Additional information was requested and the following was obtained: did device not feed clips? yes.Did device feed clips sideways? unknown.Did device not fire clips (jammed)? yes.Did device fire malformed clips? unknown.Did device fire scissored clips? yes.Did device drop or eject clips? no.Was the handle broken? no.Were the jaws broken? no.Was the shaft broken? no.An analysis of the product could not be performed since a physical sample was not received for evaluation.However, if the product is received at a later date, the investigation will be updated as applicable.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformance were identified.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.A manufacturing record evaluation was performed for the finished device lot number, and no [non-conformances / manufacturing irregularities] were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIGAMAX-5MM ENDO CLIP APPLIER
Type of Device
CLIP, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*  00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*   00969
Manufacturer Contact
kate karberg
475 calle c
guaynabo 
*  
3035526892
MDR Report Key15681412
MDR Text Key306899704
Report Number3005075853-2022-07308
Device Sequence Number1
Product Code FZP
UDI-Device Identifier10705036001843
UDI-Public10705036001843
Combination Product (y/n)N
Reporter Country CodeTW
PMA/PMN Number
K050344
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign
Reporter Occupation Other
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberEL5ML
Device Catalogue NumberEL5ML
Device Lot NumberX9506V
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 10/16/2022
Initial Date FDA Received10/27/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/03/2022
Is the Device Single Use? Yes
Type of Device Usage Initial