OAKDALE LEVEL 1 HOTLINE LOW FLOW SYSTEMS - HL-90; WARMER, THERMAL, INFUSION FLUID Back to Search Results
Model Number HL-90
Device Problem Failure to Power Up (1476)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/30/2022
Event Type  malfunction  
Manufacturer Narrative
Udi and operator of device are unknown.No information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Event Description
It was reported that the unit is not powering on.No patient involvement reported.
 
Manufacturer Narrative
Other, other text: h6: event problem and evaluation codes: updated.H10: manufacturing device history record review was not performed because the results of the complaint investigation do not indicate a problem with the manufacture of the device.No causes or potential causes of the customer's reported problem were found during the review of service and repair records.A product sample was received for evaluation.Visual and functional testing was performed.Visual inspection found cracked enclosure and tank cover.Damaged power switch and printed circuit board (pcb).Functional testing found the connection between the pcb and power switch is damaged.The root cause of the reported issue was found to be customer damage from use of the power switch.No action taken due to the age and condition of the device.It is deemed beyond economical repair and will be scrapped.This issue is being monitored via an internal capa which corrected this issue (pcb to power switch connection damage) with design improvement.If the occurrence of this issue increases significantly, additional corrective actions will be taken.This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 HOTLINE LOW FLOW SYSTEMS - HL-90
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
OAKDALE
3350 granada ave n
oakdale MN 55128
Manufacturer (Section G)
NULL
3350 granada ave n
oakdale MN 55128
Manufacturer Contact
jim vegel
6000 nathan lane north
attn: spd strm/or deliveries
minneapolis, MN 55442
MDR Report Key15681687
MDR Text Key306812029
Report Number3012307300-2022-26367
Device Sequence Number1
Product Code LGZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K001764
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 04/24/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberHL-90
Device Catalogue NumberCON-HL-90
Was Device Available for Evaluation? Device Returned to Manufacturer
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/30/2022
Initial Date FDA Received10/27/2022
Supplement Dates Manufacturer Received12/28/2022
Supplement Dates FDA Received04/24/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/25/2003
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-