BAXTER HEALTHCARE CORPORATION BURETROL SOLUTION SETS; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Catalog Number ARC7503P
Device Problems Fluid/Blood Leak (1250); Component Missing (2306)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/21/2022
Event Type  malfunction  
Event Description
It was reported that the cap of a buretrol solution set was missing and liquids leaked.This was observed prior to use.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
Initial reporter address: (b)(6).Initial reporter phone no.: (b)(6).The device was received and is currently awaiting evaluation.Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
H10: the device was received for evaluation.A visual inspection was performed which observed the absence of the filter.No leaks were detected during sample analysis; the report of leak was determined to be related to the absence of the filter.The reported condition of missing cap was not verified, however, the condition of missing filter component was verified.The cause of the condition could not be determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BURETROL SOLUTION SETS
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - CALI
calle 36 no. 2c-22,
apartado aero 2446
cali
CO  
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 
2242702068
MDR Report Key15683633
MDR Text Key305162498
Report Number1416980-2022-05788
Device Sequence Number1
Product Code FPA
UDI-Device Identifier00085412479484
UDI-Public(01)00085412479484
Combination Product (y/n)N
Reporter Country CodeCO
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/27/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberARC7503P
Device Lot NumberSE22GG1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/07/2022
Is the Reporter a Health Professional? No
Date Manufacturer Received11/26/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/12/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-