Model Number XS PST
Device Problem Image Display Error/Artifact (1304)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/03/2022
Event Type  malfunction  
Event Description
Customer is reporting a display issue with coaguchek xs meter serial number (b)(4).The customer stated the display has faded spots.Customer states the results field is not missing segments but the segments are faded.Customer did not report any misinterpretation of results.
 
Manufacturer Narrative
Occupation is patient/consumer.Meter was requested for return and the investigation is ongoing.
 
Manufacturer Narrative
The meter was returned for investigation.The device did not turn on during investigation, therefore the error log could not be read out.The optical inspection of the printed circuit board revealed contamination due to residues of fluid ingress.The contamination on the battery contacts caused a power interruption.Therefore it is not possible to turn the meter on.The contamination can lead to the mentioned problem with the faded/ missing segments.This malfunction is related to a user error.Medwatch fields d9 and h3 have been updated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COAGUCHEK ® XS SYSTEM
Type of Device
PROTHROMBIN TIME MONITOR
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
na
indianapolis, IN 46250
3174767531
MDR Report Key15683718
MDR Text Key307068997
Report Number1823260-2022-03386
Device Sequence Number1
Product Code GJS
UDI-Device Identifier00365702125100
UDI-Public00365702125100
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K062925
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/27/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberXS PST
Device Catalogue Number04837738001
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Date Manufacturer Received11/08/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient SexFemale
-
-