Catalog Number UNKNOWN
Device Problem Osseointegration Problem (3003)
Patient Problem Inadequate Osseointegration (2646)
Event Date 08/01/2022
Event Type  Injury  
Manufacturer Narrative
Internal complaint reference: (b)(4).Guntin, j., bartosiak, k.A., della valle, c.J., patel, a., gerlinger, t.L., & nam, d.(2022).Radiographic and clinical analysis of a porous-coated metaphyseal cone for revision total knee arthroplasty.The knee, 37, 162-170.Doi.Org/10.1016/j.Knee.2022.04.003.
 
Event Description
It was reported that on literature review "radiographic and clinical analysis of a porous-coated metaphyseal cone for revision total knee arthroplasty", one (1) patient suffered from femoral component loosening after having a tka index revision surgery performed using legion metaphyseal cone system.This patient underwent an unspecified additional surgical intervention.The outcome of the patient is unknown.No further information is available.
 
Manufacturer Narrative
H3, h6: the device was not returned for evaluation and the reported event could not be confirmed.The clinical/medical investigation concluded that, per complaint details, a patient suffered from femoral component loosening after having an index revision tka surgery (without utilization of the femoral cone) and underwent an unspecified additional surgical intervention.The outcome of the patient is unknown.It was communicated that no further information is available.Without the requested clinical documentation, no clinical factors could be definitively concluded to have contributed to the reported events.The images provided in the article have been interpreted within the text; therefore, no further analysis of the images is required.The patient impact beyond that which was reported could not be determined.No further medical assessment can be rendered at this time.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Factors that could contribute to the reported event include abnormal motion over time, bone degeneration, size selected, loss of ingrowth, osteolysis and/or traumatic injury.The contribution of the device to the reported event could not be corroborated.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN LEGION TOTAL KNEE REV FEM COMP
Type of Device
PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key15684838
MDR Text Key302563944
Report Number1020279-2022-04597
Device Sequence Number1
Product Code KRO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K151118
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Study,Literature,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue NumberUNKNOWN
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/04/2022
Initial Date FDA Received10/27/2022
Supplement Dates Manufacturer Received11/16/2022
Supplement Dates FDA Received11/21/2022
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-