MAUDE Adverse Event Report: MAQUET GMBH FIXTURE FOR BODY SUPPORTS; TABLE AND ATTACHMENTS, OPERATING-ROOM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET GMBH FIXTURE FOR BODY SUPPORTS; TABLE AND ATTACHMENTS, OPERATING-ROOM Back to Search Results
Model Number 100219C0
Device Problems Use of Device Problem (1670); Positioning Problem (3009)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Additional information will be provided following the conclusion of the investigation.Device not returned to manufacturer.
 
Event Description
On 25th october, 2022 getinge became aware of an issue with one of accessories ¿ 100219c0 - fixture for body supports.As it was stated, the patient partially fell off the operating table during positioning.The patient did not hit ground or any object.The staff adjusted patient's position and the operation continued without complications.There was no injury reported, however, we decided to report the issue in abundance of caution based on the potential the similar event could lead to serious injury in case of event recurrence.
 
Manufacturer Narrative
Getinge became aware of an issue with one of the accessories ¿ 100219c0 - fixture for body supports used with 113112b0 betastar operating table.The patient partially fell off the operating table during positioning.The patient did not hit the ground or any object.The staff adjusted patient's position and the operation continued without complications.There was no injury reported, however, we decided to report the issue in abundance of caution based on the potential for serious injury if the situation, namely patient partially falling off the table, was to reoccur.The affected getinge device has been evaluated by the company¿s service technician.The technician observed that the joints must have been under-tightened, as all clamping joints were working properly.According to all available information, no malfunction with the fixture for body supports could be found.In the instruction for use (ga 1002.19 en 11, page 5) there is information that the fixture 100219c0 is designed to be used to mount certain body supports to secure the patient.Any use other than that is not in compliance with the intended purpose.In the ifu (ga 1002.19 en 11, page 6) the user is warned about the risk coming from improper handling, incorrect positioning and unauthorized modifications of the device.The user is also warned that if a product or accessory is not fixed properly it may become loose and lead to injury (ga 1002.19 en 11, page 7).As it is very likely that the user utilized the accessory disregarding safety notes and suggestions from the user manual, the root cause for this issue is user error.The sales representative has trained the staff how to use the device properly.With the investigation performed it was concluded that upon the event occurrence, the device was being used for the patient¿s treatment, thus was also directly involved with the reported incident.As no malfunction with the fixture for body supports could be found, it was considered that the getinge device was up to the specification.In summary, this complaint is the first one registered on maquet gmbh fixture for body supports with catalog number 100219c0 where the patient partially fell off the table.Comparing the number of complained devices (1) to the number of units placed on the market (60258) we can conclude the failure ratio is 0,002% for the issue investigated herein and that this is a single and isolated case.We currently do not have any information that would warrant further action towards the device manufacturing or devices on the market, however as per our complaint handling processes will continue to monitor the customer experiences with the device for any future information.The correction of d1 brand name, b5 describe event or problem, h8 usage of device fields deems required.This is based on the internal evaluation.Previous d1 brand name: accessories universal corrected d1 brand name: fixture for body supports previous b5 describe event or problem: on 25th october, 2022 getinge became aware of an issue with one of accessories ¿ 100219c0 - fixture for body supports.As it was stated, the patient partially fell off the operating table during positioning.The patient did not hit ground or any object.The staff adjusted patient's position and the operation continued without complications.There was no injury reported, however, we decided to report the issue in abundance of caution based on the potential the similar event could lead to serious injury in case of event recurrence.Corrected b5 describe event or problem: on 25th october 2022 getinge became aware of an issue with one of the accessories ¿ 100219c0 - fixture for body supports used with 113112b0 betastar operating table.The patient partially fell off the operating table during positioning.The patient did not hit the ground or any object.The staff adjusted patient's position and the operation continued without complications.There was no injury reported, however, we decided to report the issue in abundance of caution based on the potential for serious injury if the situation, namely patient partially falling off the table, was to reoccur.Previous h8 usage of device: unknown corrected h8 usage of device: reuse.
 
Event Description
On 25th october 2022 getinge became aware of an issue with one of the accessories ¿ 100219c0 - fixture for body supports used with 113112b0 betastar operating table.The patient partially fell off the operating table during positioning.The patient did not hit the ground or any object.The staff adjusted patient's position and the operation continued without complications.There was no injury reported, however, we decided to report the issue in abundance of caution based on the potential for serious injury if the situation, namely patient partially falling off the table, was to reoccur.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FIXTURE FOR BODY SUPPORTS
Type of Device
TABLE AND ATTACHMENTS, OPERATING-ROOM
Manufacturer (Section D)
MAQUET GMBH
kehler strasse 31
rastatt
GM 
Manufacturer (Section G)
MAQUET GMBH
kehler strasse 31
rastatt
GM  
Manufacturer Contact
holger ullrich
kehler strasse 31
rastatt 
GM  
MDR Report Key15689759
MDR Text Key303652748
Report Number8010652-2022-00037
Device Sequence Number1
Product Code BWN
Combination Product (y/n)N
Reporter Country CodeEZ
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 10/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number100219C0
Device Catalogue Number100219C0
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/25/2022
Initial Date FDA Received10/28/2022
Supplement Dates Manufacturer Received02/02/2023
Supplement Dates FDA Received02/02/2023
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Type of Device Usage Reuse