• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANGIOMED GMBH & CO. MEDIZINTECHNIK KG LIFESTENT XL VASCULAR STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANGIOMED GMBH & CO. MEDIZINTECHNIK KG LIFESTENT XL VASCULAR STENT Back to Search Results
Catalog Number EX061203C
Device Problems Break (1069); Difficult to Remove (1528); Material Deformation (2976); Material Twisted/Bent (2981)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/22/2022
Event Type  malfunction  
Manufacturer Narrative
The catalog number identified has not been cleared in the us but is similar to the lifestent vascular stent products that are cleared in the us.The pro code and 510 k number for the lifestent vascular stent products are identified.As the lot number for the device was provided, a review of the device history records is currently being performed.The return of the sample is pending.However, photos and images were provided for review.The investigation of the reported event is currently underway.Device expiry date: 01/2023.
 
Event Description
It was reported that during a stent placement procedure in superficial femoral artery via left lower limb, device was allegedly fractured.The procedure was completed by using another device.There was no reported patient injury.
 
Manufacturer Narrative
H10: the catalog number identified in section d4 has not been cleared in the us but is similar to the lifestent vascular stent products that are cleared in the us.The pro code and 510 k number for the lifestent vascular stent products are identified in d2 and g4.H10: manufacturing review: the lot history records of this lot were reviewed with special attention to the manufacturing and inspection of this product and the product was found to have met the specification prior to shipment.Based on the information available it is not reasonably suggested that a manufacturing process may have caused or contributed to the reported issue.Investigation summary: the sample was returned for evaluation.The returned sample was found without stent that reportedly had been deployed; the proximal end of the oval tip was found with deformation that may indicate that the system was stuck on the stent upon removal.Provided images have a poor resolution so that single struts are not visible and strut fracture cannot be confirmed; however, the stent contour confirms stent twist.The images do not demonstrate difficult removal.The investigation leads to confirmed result for stent twist.The vessel was not tortuous/ calcified, the lesion was pre dilated, and the user did not experience difficulty during deployment.Based on the investigation of the provided information, the investigation is closed as confirmed for stent twisting.A definitive root cause for the reported event could not be determined based upon the available information.Labeling review: in reviewing the relevant labeling for this product, the potential issue was found addressed.The instructions for use closely describes holding and handling of the system during deployment; in particular the instructions for use state: 'confirm that the introducer sheath is secure and will not move during deployment.(¿) to ensure the most accurate placement, firmly hold the black system stability sheath throughout deployment.(¿) do not hold the silver stent delivery sheath at any time during deployment.Do not constrict the stent delivery sheath during stent deployment.Initiate stent deployment by rotating the thumbwheel in the direction of the arrows, while holding the handle in a fixed position.' regarding pta the ifu state: 'predilation of the lesion should be performed using standard techniques.' regarding access and accessories, the instructions for use states: 'gain femoral access utilizing a 6f (2.0 mm) or larger introducer sheath.Insert a 0.035inch (0.89 mm) diameter guidewire of appropriate length (.)'.H10: b5, d4 (expiry date: 01/2023), g3, h6 (device) h11: h6 (method, result, conclusion) h11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : see h10.
 
Event Description
It was reported that during a stent placement procedure in superficial femoral artery via left femoral artery, device was allegedly fractured.It was further reported that stent was allegedly difficult to be removed.Reportedly the stent was removed and the procedure was completed by using another device.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFESTENT XL VASCULAR STENT
Type of Device
VASCULAR STENT
Manufacturer (Section D)
ANGIOMED GMBH & CO. MEDIZINTECHNIK KG
wachhausstr. 6
karlsruhe 76227
GM  76227
Manufacturer (Section G)
ANGIOMED GMBH & CO. MEDIZINTECHNIK KG
wachhausstr. 6
karlsruhe 76227
GM   76227
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key15690831
MDR Text Key306999595
Report Number9681442-2022-00316
Device Sequence Number1
Product Code NIP
UDI-Device Identifier04049519001463
UDI-Public(01)04049519001463
Combination Product (y/n)N
Reporter Country CodeBR
PMA/PMN Number
P070014
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative,Distributor
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 01/17/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/28/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberEX061203C
Device Lot NumberANFN3737
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Manufacturer Received01/20/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-