• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: KARL STORZ SE & CO. KG CUTTING LOOP, BIPOLAR, 24/26FR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

KARL STORZ SE & CO. KG CUTTING LOOP, BIPOLAR, 24/26FR Back to Search Results
Model Number 27040GP1
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/13/2022
Event Type  malfunction  
Manufacturer Narrative
The affected device has been requested for investigation by the manufacturer.Device was not yet returned for investigation.
 
Event Description
It was reported that during a hystero resection, the loop broke 3 cm from its end as well as the wire.The ruptured part remained in the endometrium and required removal with hysteroscope forceps.No confirmation of injury or any adverse impact to the patient, user or third party reported.The broken part was retrieved with hysteroscope forceps.Internal karl storz reference number: (b)(4).
 
Manufacturer Narrative
Based on the description of the defect in conjunction with complaint data, the cutting loop has already been used several times and the loop was probably worn.In addition, increased force on the loop may also have caused the sling to break.Therefore, the most probably root cause is mechanical overload due to excessive force.The event is filed under internal karl storz complaint id: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CUTTING LOOP, BIPOLAR, 24/26FR
Type of Device
CUTTING LOOP
Manufacturer (Section D)
KARL STORZ SE & CO. KG
dr.-karl-storz-strasse 34
tuttlingen, 78532
GM  78532
Manufacturer (Section G)
KARL STORZ SE & CO. KG
dr.-karl-storz-strasse 34
tuttlingen, 78532
GM   78532
Manufacturer Contact
anja fair
2151 e grand ave
el segundo, CA 90245-5017
4242188738
MDR Report Key15693894
MDR Text Key306988389
Report Number9610617-2022-00297
Device Sequence Number1
Product Code FAS
UDI-Device Identifier04048551393482
UDI-Public4048551393482
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K061541
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 01/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/28/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number27040GP1
Device Catalogue Number27040GP1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/19/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-