OAKDALE LEVEL 1 HOTLINE FLUID WARMER; WARMER, THERMAL, INFUSION FLUID Back to Search Results
Catalog Number CON-HL-90
Device Problems Display or Visual Feedback Problem (1184); Noise, Audible (3273)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
This mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).(udi) was unknown.No product information has been provided to date.The original device history record (dhr) is no longer applicable as device was manufactured in 1997.No previous service history was found.A product sample was received for evaluation.Visual and functional testing were performed.Visual inspection showed display was damaged.During functional testing using the reported issue was duplicated.The root cause of the problem was related to normal wear and tear of the device.No action will be taken as the device will be scrapped due to age.
 
Event Description
It was reported that the device's pump was noisy and the display was broken.No patient injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 HOTLINE FLUID WARMER
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
OAKDALE
3350 granada ave n
oakdale MN 55128
Manufacturer (Section G)
OAKDALE
3350 granada ave n
oakdale MN 55128
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key15694534
MDR Text Key306885820
Report Number3012307300-2022-26440
Device Sequence Number1
Product Code LGZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K001764
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial
Report Date 10/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/29/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberCON-HL-90
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/15/2019
Was the Report Sent to FDA? No
Date Manufacturer Received10/07/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/25/1997
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No