Model Number FT4 G3
Device Problem High Test Results (2457)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/03/2022
Event Type  malfunction  
Manufacturer Narrative
The investigation is ongoing.
 
Event Description
There was an allegation of questionable elecsys ft4 iii results for 1 patient sample on a cobas e 801 module compared to an e411 module and an abbott architect.It is not known if any questionable results were reported outside of the laboratory.The customer's e801 module serial number was requested but not provided.The customer's ft4 iii reagent lot and expiration lot were requested but not provided.The investigational e401 module's serial number is (b)(4).This medwatch will cover ft4 iii.Refer to medwatch with patient identifier (b)(6) for information on the ft3 iii results and patient identifier (b)(6) for information on the ft3 iii ver.2 results.
 
Manufacturer Narrative
The investigation determined that the ft4 result differences generated between the different methods are consistent with methodological differences.The investigation did not identify a product problem.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECSYS FT4 III
Type of Device
RADIOIMMUNOASSAY, FREE THYROXINE
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
na
indianapolis, IN 46250
3174767531
MDR Report Key15696932
MDR Text Key306914070
Report Number1823260-2022-03409
Device Sequence Number1
Product Code CEC
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K131244
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 01/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/31/2022
Device Model NumberFT4 G3
Device Catalogue Number07976836190
Device Lot Number593219
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/05/2022
Initial Date FDA Received10/29/2022
Supplement Dates Manufacturer Received12/21/2022
Supplement Dates FDA Received01/13/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-