H10: manufacturing review: the lot history records of this lot were reviewed with special attention to the manufacturing and inspection of this product and the product was found to have met the specification prior to shipment.Investigation summary: the stent delivery system was returned for evaluation.The stent was found partially deployed upon receipt.A force transmitting sheath inside the system was found broken, which made a successful deployment impossible.Device compatible accessories were used.Based on the information available, breakage of a force transmitting component is confirmed, which is related to the reported impossibility for complete stent deployment.A definitive root cause could not be determined based upon the available information.Labeling review: in reviewing the relevant labeling for this product, the potential issue was found addressed.Based on the instructions for use supplied with this product delivery system specific events that could be associated with clinical complications include but are not limited to failure to deploy and high deployment forces.Regarding correct deployment the instructions for use states: "maintain a stationary hold on the white stability sheath during covered stent deployment.Hold the white stability sheath as close as possible to the introducer without touching the dark brown moving catheter of the distal catheter assembly.Maintain the remainder of the white stability sheath relaxed and avoid tension." regarding preparation the instructions for use states: "pre-dilate the stenosis with a pta balloon catheter of appropriate length and diameter for the lesion to be treated." h10: d4 (expiration date: 01/2024).H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|