Model Number V60 VENT, CHINA OPT: CFLEX, AVAPS
Device Problem Volume Accuracy Problem (1675)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/08/2022
Event Type  malfunction  
Manufacturer Narrative
Reporting address state: (b)(6).Reporting address postal: (b)(6).Reporting institution phone #: (b)(6).Reporter phone #: (b)(6).
 
Event Description
Philips received a complaint on the v60 ventilator, indicating that the unit alarmed: low tidal volume.The device was in clinical use at the time the issue was discovered, however, there was no patient or user harm as a result.The philips authorized service provider (asp) evaluated the device and reported the unit alarmed appropriately in response to the condition.There was no delay in patient therapy.The unit was exchanged for a alternative unit.The asp reported there has been no repair.The customer elected to discontinue service and remove the unit from use in lieu of repair.Therefore, the investigation concludes that no further action is required at this time.If additional information is received the complaint file will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESPIRONICS
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
Manufacturer (Section D)
RESPIRONICS CALIFORNIA, LLC
2271 cosmos court
carlsbad CA 92011
Manufacturer (Section G)
RESPIRONICS CALIFORNIA, LLC
2271 cosmos court
carlsbad CA 92011
Manufacturer Contact
kimberly shelly
2271 cosmos court
carlsbad, CA 92011
7609187300
MDR Report Key15709267
MDR Text Key306496231
Report Number2031642-2022-02687
Device Sequence Number1
Product Code MNT
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K102985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Other
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberV60 VENT, CHINA OPT: CFLEX, AVAPS
Device Catalogue Number1076716
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 10/08/2022
Initial Date FDA Received11/01/2022
Date Device Manufactured06/05/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-