AOMORI OLYMPUS CO., LTD. SINGLE USE LIGATING DEVICE Back to Search Results
Model Number HX-400U-30
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/05/2022
Event Type  malfunction  
Event Description
The customer reported to olympus that while using the single use ligating device, the loop could not be released.The patient cut the sheath to remove the scope and removed the loop by cutting the loop with a loop cutter.The therapeutic procedure (colon polypectomy) was completed using a similar device, and there was no patient harm associated with the event.
 
Manufacturer Narrative
The device was returned and evaluated, and the customer¿s allegation was not confirmed.The operation pipe of the operation part was broken.The tube sheath of the insertion part was cut at a position about 1930mm from the insertion part side.The coil sheath was cut at a position approximately 1850mm from the tip of the insertion tube.The investigation is ongoing and follow-up with the user facility is currently being performed.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the condition of the subject device and similar cases in the past, a likely factor causing the reported phenomenon might be the following: - the proximate side of the loop was temporarily retracted into the coil sheath, and the loop was caught in between the coil sheath and the hook.This prevented the slider from moving and the loop could not be detached from the coil sheath.Under these circumstances, the slider was forcefully operated.As a result, the operating pipe was bent and broken.The phenomenon described above was likely caused by the mechanism below: 1) the loop was detached from the hook while the coil sheath was being retracted into the tube sheath, or the loop was hung on to the tissue and it was positionally fixed at the distal end of the tube.As a result, the loop was unable to release.2) the tube sheath was pulled toward the proximate side while the hook was extended from the distal end of the coil sheath.3) since the tube sheath was pulled toward the proximate side, causing the tube to pull the loop.This caused the loop to retract into the coil sheath.As a result, the loop was caught in between the hook and inner surface of the coil, and the loop stopped moving.4) the slider was forcefully operated in state of 3) description causing the operation pipe to break.The severed device of distal side was not returned for investigation; therefore, the reported event could not be confirmed.The event can be prevented by following the instructions for use which state: - "do not strike or crush the coil sheath during operation.Doing so can damage the distal end of the coil sheath, which could make it impossible to detach the loop after ligation.In this case, refer to section 12, ¿emergency treatment¿ and as shown ¿equipment to be used in an emergency¿ on page 3 in this manual." - "do not remove the loop from the hook while the coil sheath is not extended from the tube sheath.Otherwise, the loop may be tangled with the hook and become impossible to be removed.In this case, refer to section 12, ¿emergency treatment¿ and as shown ¿equipment to be used in an emergency¿ on page 3 in this manual." - "do not hold the loop with the distal end of the tube sheath while the loop is surrounding the tissue.Otherwise, when the tissue is ligated, the loop may be detached from the hook in the tube sheath and tangled with the hook.That may make the loop impossible to be removed.In this case, refer to section 12, ¿emergency treatment¿ and as shown ¿equipment to be used in an emergency¿ on page 3 in this manual." - "never use excessive force to operate the instrument.This could damage the instrument." olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SINGLE USE LIGATING DEVICE
Type of Device
SINGLE USE LIGATING DEVICE
Manufacturer (Section D)
AOMORI OLYMPUS CO., LTD.
2-248-1 okkonoki
kuroishi-shi, aomori 036-0 357
JA  036-0357
Manufacturer Contact
masaharu hirose
2-248-1 okkonoki
kuroishi-shi, aomori 036-0-357
JA   036-0357
426422891
MDR Report Key15710222
MDR Text Key307106936
Report Number9614641-2022-00556
Device Sequence Number1
Product Code FHN
UDI-Device Identifier04953170368615
UDI-Public04953170368615
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
CLASS2-EXMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 11/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/01/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberHX-400U-30
Device Lot Number22K
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/12/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received11/08/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/13/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-