Model Number 58K16906
Device Problem Detachment of Device or Device Component (2907)
Patient Problem Hemorrhage/Bleeding (1888)
Event Date 10/07/2022
Event Type  malfunction  
Manufacturer Narrative
The device evaluation is anticipated.However, the complaint cannot be confirmed without the completion of a product evaluation.A supplemental report will be forthcoming with the evaluation.The lot number was not provided thus a device history record was not reviewed.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis and any excursions above the control limits are assessed and documented as a part of the monthly review.
 
Event Description
It was reported that the pressure tubing got detached at the connection with the three way stopcock, which caused blood leakage on the first day of use.The event occurred to an adult patient and although the exact amount of blood leakage was unknown, neither any additional treatments nor extension of hospitalization was needed.The device was exchanged and the problem was solved.Although there was blood leakage observed, there were no patient complications reported by the customer.
 
Manufacturer Narrative
Additional testing was completed.The tubing od, 0.1110" near the point of detachment, was within specification.Stopcock tubing housing id was not able to be measured due to presence of bonding material on the inner wall.Further evaluation regarding related quality issues is under investigation.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis and any excursions above the control limits are assessed and documented as a part of the monthly review.
 
Manufacturer Narrative
The reported event of pressure tubing got detached at the connection with the three way stopcock was confirmed.As received, pressure tubing was detached from three way stopcock.Bond indication was observed on inserted portion of the detached tubing.Further evaluation regarding related quality issues is under investigation.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis and any excursions above the control limits are assessed and documented as a part of the monthly review.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRESSURE TUBING WITH 3 WAY STOPCOCK
Type of Device
TRANSDUCER, BLOOD PRESSURE, EXTRAVASCULAR
Manufacturer (Section D)
EDWARDS LIFESCIENCES DR
parque industrial de itabo
carr. sanchez km 18.5
haina, san cristobal
Manufacturer (Section G)
EDWARDS LIFESCIENCES PR
state rd indus pk 402 km 1.4
anasco
Manufacturer Contact
jessica atallah
1 edwards way
irvine, CA 92614
MDR Report Key15710523
MDR Text Key304824763
Report Number2015691-2022-08939
Device Sequence Number1
Product Code DRS
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K832907
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date 12/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number58K16906
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/26/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/11/2022
Initial Date FDA Received11/01/2022
Supplement Dates Manufacturer Received11/07/2022
12/01/2022
Supplement Dates FDA Received11/23/2022
12/19/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-