OAKDALE LEVEL 1 HOTLINE LOW FLOW SYSTEMS - HL-90; WARMER, THERMAL, INFUSION FLUID Back to Search Results
Model Number HL-90
Device Problem Noise, Audible (3273)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Udi section is unknown, no information has been provided to date.Device evaluation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Event Description
It was reported the device's motor was real loud.There has been no report of patient involvement or no observable clinical symptoms or a change in symptoms identified in the patient.
 
Manufacturer Narrative
H10: device evaluation: the device was returned for investigation.A visual inspection and functional test were performed.Visual inspection found the device presented with stained reservoir, faded line cord, worn enclosure and front cover, and cracked tank cover.The customer stated problem was duplicated.The motor on the pump was really loud when running due to worn out pump.Due to wear of age, the water pump was overdue for replacement after 17 years of usage.As a result, a fault could not be assigned.No action was taken due to the age and condition of the device.It was deemed beyond economical repair and it was scrapped.The cause of the reported problem was traced to the wear/usage of the device.A dhr (device history review) was not performed because the device is beyond a year from its manufacture date and there was no indication of a manufacturing defect during the investigation.Service history review identified this device has not been in for service previously.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 HOTLINE LOW FLOW SYSTEMS - HL-90
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
OAKDALE
3350 granada ave n
oakdale MN 55128
Manufacturer (Section G)
NULL
3350 granada ave n
oakdale MN 55128
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key15710731
MDR Text Key307169726
Report Number3012307300-2022-26590
Device Sequence Number1
Product Code LGZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K001764
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberHL-90
Device Catalogue NumberCON-HL-90
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/17/2022
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/05/2022
Initial Date FDA Received11/01/2022
Supplement Dates Manufacturer Received11/23/2022
Supplement Dates FDA Received12/02/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/13/2005
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-