Catalog Number 5C4482
Device Problems Difficult to Open or Close (2921); Material Integrity Problem (2978)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/06/2022
Event Type  malfunction  
Manufacturer Narrative
Initial reporter address: (b)(6).The device was received and is currently awaiting evaluation.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that the twist clamp (white sleeve) of a minicap transfer set was damaged; further described as ¿no flute on white sleeve¿ resulting in ¿a gap after twisting¿ (gap between sleeve and mainbody).This was observed during before use of the device for peritoneal dialysis therapy.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
H10: the device was received for evaluation.A visual inspection with the naked eye observed the lock notch on both sides of the twist clamp was not present.The reported condition was verified.The direct cause was determined to be related to a manufacturing issue.This condition would not result in a breach of the sterile pathway; therefore, would not lead to a serious injury.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MINICAP TRANSFER SET
Type of Device
SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - MOUNTAIN HOME
1900 n highway 201
mountain home AR 72653
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 
2242702068
MDR Report Key15711982
MDR Text Key304632107
Report Number1416980-2022-05951
Device Sequence Number1
Product Code KDJ
Combination Product (y/n)N
Reporter Country CodeTW
PMA/PMN Number
K152675
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 12/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number5C4482
Device Lot NumberH22A04061
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/25/2022
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/06/2022
Initial Date FDA Received11/01/2022
Supplement Dates Manufacturer Received11/28/2022
Supplement Dates FDA Received12/07/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-