The reported event is confirmed cause unknown.Visual evaluation noted 1 statlock stabilization device received with no original packaging.Swivel base and clamp are detached from pad.Therefore, it is unknown how the product meets specifications.Although an exact root cause could not be determined a potential root cause could be glue selection incompatible with clamp base.The lot number was unknown; therefore, the device history record could not be reviewed.The instructions for use were found adequate and state the following: "indications for use: the statlock® device is a stabilization device for compatible catheters.Contraindications: known tape or adhesive allergies.Warnings and precautions: 1.Do not use the statlock® device where loss of adherence could occur, such as with a confused patient, diaphoretic or non-adherent skin, or when the access device is not monitored daily.2.Observe universal blood and body fluid precautions and infection control procedures, during application and removal of the statlock® device.3.Minimize catheter manipulation during application and removal of the statlock® device.4.Daily maintenance: a.The statlock® device should be assessed daily and changed when clinically indicated, at least every seven days.B.If pad becomes soiled, wash with soap/water, saline or hydrogen peroxide.Do not use alcohol or prepackaged bathing systems, which could lead to early lifting.C.If showering/bathing, cover with plastic wrap or waterproof dressing.D.Conduct skin assessment prior to application and repeat daily per facility protocol.E.Use clinical judgment on the removal of the statlock® stabilization device if the patient experiences any fluid shifts that may interfere with skin integrity.The statlock® stabilization device should be monitored daily and replaced when clinically indicated, at least every 7 days.The catheter insertion site should be treated per local policy recommendations.The statlock® device is contraindicated on patients with known tape and adhesive allergies.Alcohol and acetone may weaken the adhesive bond between the statlock® device pad and the skin.Care should be taken when using these solutions while performing catheter site care.Please consult product inserts and labels for any indications, contraindications, hazards, warnings, cautions and directions for use.For additional information on the statlock® foley stabilization device, please call your local bard representative.© copyright 2019 c.R.Bard, inc.Bard and statlock are trademarks and/or registered trademarks of c.R.Bard, inc.Protected by domestic and foreign patents and patents pending.Pk2405563 03/2019 manufacturer: c.R.Bard, inc.Covington, ga 30014 usa 800-526-4455 www.Bardmedical.Com made in mexico instructions for use foley application technique prep 1.Place foley catheter into retainer.Directional arrow should point towards catheter tip, and balloon inflation arm should be next to the clamp hinge.2.Close lid, being careful to avoid pinching the catheter.3.Identify securement site by laying the device retainer on the front of the thigh, leaving 1 inch of catheter slack between insertion site and the statlock® device retainer.4.After placing the statlock® stabilization device off to the side, cleanse and degrease the securement site with alcohol per hospital policy.Let skin dry.5.Apply skin protectant, in direction of hair growth, to area larger than securement site.Allow to dry completely (10-15 seconds).6.Using permanent marker, write initials and date of application on the statlock® device anchor pad.Note: always secure catheter into the statlock® device retainer before applying adhesive pad on skin.Place and peel 7.Align the statlock® stabilization device over securement site leaving 1 inch of catheter slack.Make sure leg is fully extended.8.While holding the retainer to keep the pad in place, peel away paper backing, one side at a time and place tension-free on skin.Removal technique disengage 1.Open retainer by pressing release button with thumb, then lift to open.2.Remove foley catheter from the statlock® device.Dissolve 3.Wipe the edge of the pad using at least 5-6 alcohol pads until a corner lifts.Then continue to stroke undersurface of pad with alcohol to dissolve adhesive pad away.' h11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the actual/suspected device was inspected.
|