• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC ADVANCED ENERGY, LLC UNKNOWN MAE GENERATOR; ELECTROSURGICAL, CUTTING & COAGULATION & ACCES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC ADVANCED ENERGY, LLC UNKNOWN MAE GENERATOR; ELECTROSURGICAL, CUTTING & COAGULATION & ACCES Back to Search Results
Model Number MAE UNK GEN
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Arrhythmia (1721)
Event Date 10/07/2022
Event Type  Injury  
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Event Description
Information was received from healthcare provider (hcp) via a manufacturer representative regarding a handpiece.It was reported that during a box change, under general anesthetic, the doctor began using a handpiece to open device pocket.The patient spontaneously went into ventricular fibrillation (vf).The doctor provided chest compressions for short period and then a 200j external defibrillator shock was provided to bring patient out of arrhythmia.The patient was successfully reverted after one shock.Procedure afterwards continued without incident.Unclear cause for arrhythmia, doctors speculated possibly due to anesthetic reaction or handpiece conducting through device and lead.There was no further patient symptoms or complications associated with this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN MAE GENERATOR
Type of Device
ELECTROSURGICAL, CUTTING & COAGULATION & ACCES
Manufacturer (Section D)
MEDTRONIC ADVANCED ENERGY, LLC
180 international drive
portsmouth NH 03801
Manufacturer (Section G)
MEDTRONIC ADVANCED ENERGY, LLC
180 international drive
portsmouth NH 03801
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key15714501
MDR Text Key302844960
Report Number1226420-2022-00047
Device Sequence Number1
Product Code GEI
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
K143175
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 11/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberMAE UNK GEN
Device Catalogue NumberMAE UNK GEN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/07/2022
Initial Date FDA Received11/02/2022
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Life Threatening;
Patient SexFemale
-
-