SYNCARDIA SYSTEMS, LLC SYNCARDIA FREEDOM DRIVER; EXTERNAL PNEUMATIC DEVICE Back to Search Results
Catalog Number 595000-001
Device Problem Audible Prompt/Feedback Problem (4020)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/25/2022
Event Type  malfunction  
Manufacturer Narrative
The freedom driver will be returned to syncardia for evaluation.The results will be provided in a follow-up emdr.
 
Event Description
The customer, a syncardia authorize distributor, reported that the patient's hospital standby freedom driver exhibited a fault alarm upon start-up.The freedom driver was not supporting a patient at the time of the event.
 
Manufacturer Narrative
Device history record (dhr) review confirmed that freedom driver s/n (b)(6) was serviced and passed all functional testing prior to being released to finished goods.Alarm history and patient data file review found one new alarm indicating secondary motor engagement recorded in the driver's data file.However, as the secondary motor was found in the default position, there is no evidence secondary motor engagement occurred.The fault alarm was likely produced at syncardia during the investigation.Visual inspection of external components found no abnormalities.Visual inspection of internal components found no abnormalities.Freedom driver passed all functional testing for acceptance at incoming inspection.Additional testing was performed in an attempt to replicate the customer reported fault alarm.A power cycle test was performed, no alarms were produced.Customer complaint could not be replicated.The freedom driver passed all functional testing at incoming inspection and during investigation testing.Complaint could not be replicated.The failure investigation found no evidence of a device malfunction.Device was not in use by a patient at time of complaint.This issue will be monitored and trended as part of the customer complaint process.Syncardia has completed its evaluation and is closing this file.If new or additional information is received in the future, syncardia will file a follow-up mdr.
 
Event Description
The customer, a syncardia authorize distributor reported that the patient's hospital standby freedom driver exhibited a fault alarm upon start-up.The freedom driver was not supporting a patient at the time of the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA FREEDOM DRIVER
Type of Device
EXTERNAL PNEUMATIC DEVICE
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC
1992 e silverlake road
tucson AZ 85713
Manufacturer (Section G)
SYNCARDIA SYSTEMS, LLC
1992 e silverlake road
tucson AZ 85713
Manufacturer Contact
aaron meier
1992 e silverlake road
tucson, AZ 85713
5205451234
MDR Report Key15723760
MDR Text Key305088258
Report Number3003761017-2022-00121
Device Sequence Number1
Product Code LOZ
UDI-Device Identifier00858000003121
UDI-Public(01)00858000003121
Combination Product (y/n)N
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Distributor
Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date 10/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue Number595000-001
Was Device Available for Evaluation? Device Returned to Manufacturer
Initial Date Manufacturer Received 10/25/2022
Initial Date FDA Received11/04/2022
Supplement Dates Manufacturer Received12/30/2022
Supplement Dates FDA Received10/13/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/04/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age30 YR
Patient SexMale
-
-