Catalog Number 595000-001 |
Device Problem
Audible Prompt/Feedback Problem (4020)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/25/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
The freedom driver will be returned to syncardia for evaluation.The results will be provided in a follow-up emdr.
|
|
Event Description
|
The customer, a syncardia authorize distributor, reported that the patient's hospital standby freedom driver exhibited a fault alarm upon start-up.The freedom driver was not supporting a patient at the time of the event.
|
|
Manufacturer Narrative
|
Device history record (dhr) review confirmed that freedom driver s/n (b)(6) was serviced and passed all functional testing prior to being released to finished goods.Alarm history and patient data file review found one new alarm indicating secondary motor engagement recorded in the driver's data file.However, as the secondary motor was found in the default position, there is no evidence secondary motor engagement occurred.The fault alarm was likely produced at syncardia during the investigation.Visual inspection of external components found no abnormalities.Visual inspection of internal components found no abnormalities.Freedom driver passed all functional testing for acceptance at incoming inspection.Additional testing was performed in an attempt to replicate the customer reported fault alarm.A power cycle test was performed, no alarms were produced.Customer complaint could not be replicated.The freedom driver passed all functional testing at incoming inspection and during investigation testing.Complaint could not be replicated.The failure investigation found no evidence of a device malfunction.Device was not in use by a patient at time of complaint.This issue will be monitored and trended as part of the customer complaint process.Syncardia has completed its evaluation and is closing this file.If new or additional information is received in the future, syncardia will file a follow-up mdr.
|
|
Event Description
|
The customer, a syncardia authorize distributor reported that the patient's hospital standby freedom driver exhibited a fault alarm upon start-up.The freedom driver was not supporting a patient at the time of the event.
|
|
Search Alerts/Recalls
|