Catalog Number S313141 |
Device Problem
Physical Resistance/Sticking (4012)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/13/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Multiple mdr reports were filed for this event, please see associated reports: 0001825034-2022-02480.Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Event Description
|
It was reported that the shell inserter handle would not release from the shell after impaction.There were no consequences or impact to the patient.Attempts have been made and no further information is available.
|
|
Manufacturer Narrative
|
(b)(4).This follow-up report is being submitted to relay additional information.The following sections were updated/corrected updated: b4; b5; g3; h2; h3; h6 multiple mdr reports were filed for this event, please see associated reports: 0001825034-2022-02632.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Dhr was reviewed and no discrepancies related to the reported event were found.Root cause was unable to be determined as the necessary information to adequately investigation the reported event was not provided.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
No additional event information to report at this time.
|
|