MAUDE Adverse Event Report: MEDOS INTERNATIONAL SARL TRUESPAN MENISCAL REPAIR SYSTEM PEEK 12 DEGREE; SOFT-TISSUE ANCHOR, BIOABSORBABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SARL TRUESPAN MENISCAL REPAIR SYSTEM PEEK 12 DEGREE; SOFT-TISSUE ANCHOR, BIOABSORBABLE Back to Search Results
Model Number 228151
Device Problems Device-Device Incompatibility (2919); Noise, Audible (3273); Physical Resistance/Sticking (4012)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/19/2022
Event Type  malfunction  
Event Description
It was reported by the sales rep in chile that during a meniscal repair procedure on (b)(6) 2022, it was observed that the red trigger on the truespan meniscal repair system peek 12 degree device emitted a loud sound when pulled to deploy then fractured.Another like device was used to complete the procedure with a delay of 10 minutes.There were no adverse patient consequences reported.No additional information was provided.
 
Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Additional narrative: udi: (b)(4).Reporter is a j&j sales representative.Investigation summary: the complaint device has not been returned, therefore unavailable for a physical evaluation.However a video was provided.Upon visual inspection of the video, a broken red trigger can be observed on the device as the trigger is being moved by the customer.; therefore, this complaint can be confirmed.A manufacturing record evaluation was performed for the finished device lot number: 9l22268, and no nonconformances were identified.The video does not provide enough evidence to determine root cause.Without physical evaluation of the device reported, we cannot establish a root cause for the issue experienced by the customer.The possible root cause for the deploy failure reported could be attributed to a portion of tissue that was entered inside the applier needle causing a mechanical obstruction of the first plate when it was going to be deployed, this obstruction and the force applied to the trigger are contributive factors of the broken trigger.However, it cannot be conclusively affirmed.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
 
Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: investigation summary: the complaint device was received and evaluated; visual inspection reveals that the red trigger is broken.The plastic sleeve was removed to verify the presence of both plates, the plates were not found inside the applier needle and they were not returned.The device was opened to verify the broken part, it was found that the upper part of the red trigger is broken.A manufacturing record evaluation was performed for the finished device 9l22268 number, and no non-conformances were identified.As part of depuy synthes mitek quality process all devices are manufactured, inspected, and released to approved specifications.According with the visual inspection and the functional test result, this complaint can be confirmed.A possible root cause can be attributed to a portion of tissue that was entered inside the applier needle causing a mechanical obstruction of the first plate when it was going to be deployed; this obstruction and the force applied to the trigger are contributive factors of the broken trigger, however this cannot be conclusively determined.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to monitor additional complaint information for potential safety signals through complaint trending as part of post market surveillance.
 
Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H6: the actual device has been returned and is currently pending evaluation.Once reliability engineering evaluates the device, a supplemental medwatch report will be sent accordingly.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRUESPAN MENISCAL REPAIR SYSTEM PEEK 12 DEGREE
Type of Device
SOFT-TISSUE ANCHOR, BIOABSORBABLE
Manufacturer (Section D)
MEDOS INTERNATIONAL SARL
chemin blanc 38
le locle CH-24 00
SZ  CH-2400
Manufacturer (Section G)
MEDOS INTERNATIONAL SARL
chemin-blanc 38
le locle CH-24 00
SZ   CH-2400
Manufacturer Contact
kate karberg
325 paramount drive
raynham, MA 02767
3035526892
MDR Report Key15726559
MDR Text Key306623689
Report Number1221934-2022-03407
Device Sequence Number1
Product Code MBI
UDI-Device Identifier10886705026012
UDI-Public10886705026012
Combination Product (y/n)N
Reporter Country CodeCI
PMA/PMN Number
K153667
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/04/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number228151
Device Catalogue Number228151
Device Lot Number9L22268
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/19/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/07/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
-
-