Model Number PVF-M |
Device Problem
Difficult to Open or Close (2921)
|
Patient Problem
Insufficient Information (4580)
|
Event Date 09/28/2022 |
Event Type
Injury
|
Manufacturer Narrative
|
The manufacturing and material records for the perceval plus heart valve, model #pvf-m, s/n # (b)(4), as they pertain to the reported event, were retrieved and reviewed by the manufacturer¿s quality engineering.The results confirmed that this valve satisfied all material, visual, and performance standards required for a (model #pvf-m) perceval plus heart valve at the time of manufacture and release.The steady flow test was also performed.The images demonstrate the acceptable opened and closed leaflet performance of the perceval pvf-m sn# (b)(4).No anomalies are observed during the open/close cycle.The valve therefore meets the acceptance criteria of the steady flow test inspection as defined by the principal device function test at the time of release.A follow up report will be provided upon receipt of the additional information or device.
|
|
Event Description
|
The manufacturer was informed that pvf-m was implanted on (b)(6) 2022.When the cpb was finished, it was realized that one of the leaflets did not have motion.As such, the valve was explanted and another prothesis was implanted during the same day.
|
|
Manufacturer Narrative
|
The manufacturer was informed that no further information is available at this time.The valve was returned to the manufacturer.The visual inspection, performed on the returned prosthesis valve, did not highlight pre-existing defects.The hydrodynamic testing on the pvf-m was conducted.The effective orifice area (eoa) at 70 bpm, 5.0 l/min of cardiac output and mean backpressure of about 100 mmhg is 2.31 cm2, well above the iso 5840 minimum requirement 1.25 cm2.For a cardiac output of 5.0 l/min and mean aortic pressure of about 100 mmhg, the regurgitant fraction is 5.7 % and it is below the requirement of iso 5840 (rf% < 10%) for a prosthesis of equivalent tad.No anomalies were observed during the open/close cycle in both normotensive and hypotensive conditions (see attached pictures).Based on the performed analyses, it is possible to exclude the relationship between the reported issue and the returned device quality.The reported event cannot be explained by any factor intrinsic in the involved device because the claimed issues were not observed during the investigation carried out (i.E.Hydrodynamic test).As such, the definitive root cause of the event cannot be established.Should further information be received in the future, a follow up report will be provided.
|
|
Search Alerts/Recalls
|
|