• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL STANDARD MECHANICAL HEART VALVE; HEART-VALVE, MECHANICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL STANDARD MECHANICAL HEART VALVE; HEART-VALVE, MECHANICAL Back to Search Results
Catalog Number 21AEC-102
Device Problem Perivalvular Leak (1457)
Patient Problem Aortic Valve Insufficiency/ Regurgitation (4450)
Event Date 10/13/2022
Event Type  Injury  
Manufacturer Narrative
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event Description
It was reported that on (b)(6) 2022, a 21mm mechanical aortic valve and a 25mm sjm mechanical mitral valve were successfully implanted in a patient.On (b)(6) 2022, it was reported that the patient developed a tumor in the sub-valvular aorta.In addition, mild leakage was noted from the sewing cuff of the implanted aortic valve.Although the degree of regurgitation was not considered clinically significant, the decision was made to remove the tumor and the aortic device.The valve is planned to be replaced with a sjm regent valve.No issue was reported with the implanted mitral sjm valve.No patient consequences were reported.
 
Manufacturer Narrative
An event of mild leakage from the sewing cuff of the valve was reported.Information from the field indicates that the patient was diagnosed with a tumor in the subvalvular aorta which lead to valve explant.Also, the field indicates that the physician has not alleged any malfunction for abbott valves, and the valve was explanted because of a patient medical condition.A returned device assessment, to see if there were any anomalies with the valve could not be performed as the device was not returned for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STANDARD MECHANICAL HEART VALVE
Type of Device
HEART-VALVE, MECHANICAL
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ST. JUDE MEDICAL PUERTO RICO, INC. REG#2648612
20 b st caguas west park
caguas 00725
*   00725
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key15729031
MDR Text Key303049983
Report Number2135147-2022-01920
Device Sequence Number1
Product Code LWQ
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
P810002
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/18/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/04/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number21AEC-102
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received01/17/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/28/2008
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age72 YR
Patient SexFemale
-
-