MAUDE Adverse Event Report: ICU MEDICAL DE MEXICO, S. DE R.L. DE C.V. TRANSPAC® IV MONITORING KIT, DISPOSABLE TRANSDUCER, 03 ML SQUEEZE FLUSH DEVICE,; TRANSDUCER, BLOOD-PRESSURE, EXTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ICU MEDICAL DE MEXICO, S. DE R.L. DE C.V. TRANSPAC® IV MONITORING KIT, DISPOSABLE TRANSDUCER, 03 ML SQUEEZE FLUSH DEVICE,; TRANSDUCER, BLOOD-PRESSURE, EXTRAVASCULAR Back to Search Results
Catalog Number 886-42584-05
Device Problem Material Separation (1562)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/13/2022
Event Type  malfunction  
Manufacturer Narrative
The device is available to be returned for evaluation; however, it has not yet been received.
 
Event Description
It was reported that a transpac iv monitoring kit tubing was found separated once the package was opened.There were no other defects noted on the product.The product was replaced and therapy resumed.There was no patient involvement and no patient harm.No further information is available.
 
Manufacturer Narrative
The reported complaint of tubing separation was confirmed on the returned set.During visual inspection, the 12" pressure tubing was received and separated from the female luer.Insufficient adhesive coverage was observed on the tubing pocket.The separation of the bond was due to insufficient adhesive coverage on the pressure tubing.The probable cause of insufficient adhesive coverage on the pressure tubing had occurred due to an error during assembly.Additional information - d9 - date returned to mfg 11/7/2022.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRANSPAC® IV MONITORING KIT, DISPOSABLE TRANSDUCER, 03 ML SQUEEZE FLUSH DEVICE,
Type of Device
TRANSDUCER, BLOOD-PRESSURE, EXTRAVASCULAR
Manufacturer (Section D)
ICU MEDICAL DE MEXICO, S. DE R.L. DE C.V.
avenida cuarzo no. 250
ensenada, b.cfa. 22790
MX  22790
Manufacturer Contact
michael visocnik
600 n. field dr.
lake forest, IL 60045
2247062300
MDR Report Key15731448
MDR Text Key303435499
Report Number9617594-2022-00295
Device Sequence Number1
Product Code DRS
Combination Product (y/n)N
Reporter Country CodeTW
PMA/PMN Number
K061573
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative,Distributor
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 10/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number886-42584-05
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/14/2022
Initial Date FDA Received11/04/2022
Supplement Dates Manufacturer Received12/06/2022
Supplement Dates FDA Received12/08/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-