|
Catalog Number 319.100 |
Device Problem
Mechanical Jam (2983)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/01/2022 |
Event Type
malfunction
|
Event Description
|
Device report from synthes reports an event in (b)(6) as follows: it was reported that on an unknown date, the customer reported a noticeable number of issues with various length measuring devices during the last weeks.Despite correct reprocessing and care, the instruments ""hook"" function, does not run smoothly and thus complicate the surgeries or extend the surgery time because new trays must be opened.There has been no patient harm.This report involves one (1) depth gauge for large screws.This is report 2 of 6 for (b)(6).
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Event date: only event month and year are known.Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.Occupation: reporter is a j&j employee.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H3, h4, h6: the product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that the depth gauge f/scr ø4.5 - 6.5 meas-range was observed assembled and the tip of the needle is slightly bend in different directions.The observed condition was consistent as an end of life indicator for the device.No other issues were identified.A functional test was performed to device and the slider can be moved forward and backward, but it is rough running.The complaint condition was able to be replicated.After a visual inspection, it was determined that the reusable instrument device was bent from repeated use and servicing; therefore, further investigation for the reported complaint device is not required.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed conditions of depth gauge f/scr ø4.5 - 6.5 meas-range would have contributed to the complained issue.There is no indication that a design or manufacturing issue has caused the complaint condition.It was determined that the reusable instrument was worn from repeated use and servicing.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Drawing/specifications reviewed? yes, reviewed.Device history lot part # 319.100 lot # 2368795 manufacturing site: werk selzach logistik release to warehouse date: 10 jun 2008 a manufacturing record evaluation was performed for the finished article lot and no non-conformances were identified.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Search Alerts/Recalls
|
|
|