• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OAKDALE LEVEL 1 HOTLINE LOW FLOW SYSTEMS - HL-90; WARMER, THERMAL, INFUSION FLUID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OAKDALE LEVEL 1 HOTLINE LOW FLOW SYSTEMS - HL-90; WARMER, THERMAL, INFUSION FLUID Back to Search Results
Model Number HL-90
Device Problem Defective Component (2292)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/04/2022
Event Type  malfunction  
Event Description
It was reported that the unit had a broken power cord.There has been no report of patient involvement or no observable clinical symptoms or a change in symptoms identified in the patient.
 
Manufacturer Narrative
Udi is unknown.No information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Manufacturer Narrative
B3: event date updated.B5: additional information.H6: event problem and evaluation codes: updates not required.H10: device evaluation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Event Description
Additional information was received via email on (b)(6) 2022: event date updated: (b)(6) 2022.The reported problem was discovered during scheduled maintenance.No patient involvement was reported.
 
Manufacturer Narrative
Manufacturing device history record review was not performed because the results of the complaint investigation do not indicate a problem with the manufacture of the device.No causes or potential causes of the customer's reported problem were found during the review of service and repair records.A product sample was received for evaluation.Visual inspection found the prongs on the end of the line cord are bent.The root cause of the reported issue was found to be customer damage.No action taken due to the age and condition of the device.It is deemed beyond economical repair and will be scrapped.This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 HOTLINE LOW FLOW SYSTEMS - HL-90
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
OAKDALE
3350 granada ave n
oakdale MN 55128
Manufacturer (Section G)
NULL
3350 granada ave n
oakdale MN 55128
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key15734675
MDR Text Key306988980
Report Number3012307300-2022-26755
Device Sequence Number1
Product Code LGZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K001764
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date 05/25/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/04/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberHL-90
Device Catalogue NumberCON-HL-90
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/31/2022
Was the Report Sent to FDA? No
Date Manufacturer Received12/28/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/07/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-