• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER INSTRUMENTS-A DIVISION OF STRYKER CORP ZYPHR DISP CRAN PERF LRG 14/11; DRILLS, BURRS, TREPHINES & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER INSTRUMENTS-A DIVISION OF STRYKER CORP ZYPHR DISP CRAN PERF LRG 14/11; DRILLS, BURRS, TREPHINES & ACCESSORIES Back to Search Results
Model Number 5100060001
Device Problem Failure to Auto Stop (2938)
Patient Problem Hemorrhage/Bleeding (1888)
Event Date 10/07/2022
Event Type  Injury  
Event Description
It was reported that during a cranial procedure, the perforator bit did not stop when passing the dura.This lead to massive blood loss that was stopped immediately.It was further reported that no medical intervention was required.Procedure completed successfully and no post operative complications to the patient.
 
Manufacturer Narrative
A follow up report will be filed once the quality investigation is complete.
 
Manufacturer Narrative
H6: the quality investigation is complete.
 
Event Description
It was reported that during a cranial procedure, the perforator bit did not stop when passing the dura.This lead to massive blood loss that was stopped immediately.It was also reported that no further medical intervention was required.Procedure completed successfully and no post operative complications to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZYPHR DISP CRAN PERF LRG 14/11
Type of Device
DRILLS, BURRS, TREPHINES & ACCESSORIES
Manufacturer (Section D)
STRYKER INSTRUMENTS-A DIVISION OF STRYKER CORP
1941 stryker way
portage MI 49002
Manufacturer (Section G)
STRYKER INSTRUMENTS-IRELAND
instruments division
carrigtwohill bus. & tech park
carrigtwohill NA
EI   NA
Manufacturer Contact
una barry
instruments division
carrigtwohill bus. & tech park
carrigtwohill NA
EI   NA
214532900
MDR Report Key15735637
MDR Text Key303083020
Report Number3015967359-2022-02239
Device Sequence Number1
Product Code HBF
UDI-Device Identifier04546540716224
UDI-Public04546540716224
Combination Product (y/n)N
Reporter Country CodeTW
PMA/PMN Number
K082010
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Distributor
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/04/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/01/2024
Device Model Number5100060001
Device Catalogue Number5100060001
Device Lot Number21124027
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Date Manufacturer Received10/07/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/04/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-