Model Number 72081-01 |
Device Problem
Material Twisted/Bent (2981)
|
Patient Problem
Insufficient Information (4580)
|
Event Date 10/05/2022 |
Event Type
Injury
|
Manufacturer Narrative
|
At this time product has not yet been returned.An extended investigation has been performed for the reported complaint.There was no indication that the product did not meet specification.Dhrs (device history review) for the freestyle libre sensor and freestyle libre sensor kit were reviewed and the dhrs showed the freestyle libre sensor and freestyle libre sensor kit passed all tests prior to release.Shelf life studies and product monitoring & dose audits were performed during product development and on market performance continues to be monitored via product monitoring/dose audits as a part of on market performance monitoring process.If the product is returned, a physical investigation will be performed and a follow-up report submitted.All pertinent information available to abbott diabetes care has been submitted.
|
|
Event Description
|
A customer reported that the sensor tip of the adc device was bent.Due to this, the customer indicated they received unspecified third-party treatment.No additional information was provided.There was no report of death or permanent injury associated with this event.
|
|
Event Description
|
A customer reported that the sensor tip of the adc device was bent.Due to this, the customer indicated they received unspecified third-party treatment.No additional information was provided.There was no report of death or permanent injury associated with this event.
|
|
Manufacturer Narrative
|
Sensor (b)(6) has been returned and investigated.No physical damage was observed on the returned sensor patch and no issues were observed.Visual inspection was performed on the returned sensor patch.Did not observed any watermark at the base of the sensor tail.If additional product is returned, an investigation will be performed.All pertinent information available to abbott diabetes care has been submitted.
|
|