• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET SPINE INC. OPEN IMPLANT CLOSURE TOP, 5.5MM, TI; SEQUOIA PEDICLE SCREW SYSTEM INCLUDING SPEEDLINK II TRANSVERSE CONNECTOR SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET SPINE INC. OPEN IMPLANT CLOSURE TOP, 5.5MM, TI; SEQUOIA PEDICLE SCREW SYSTEM INCLUDING SPEEDLINK II TRANSVERSE CONNECTOR SYSTEM Back to Search Results
Catalog Number 3301-1
Device Problem Material Deformation (2976)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/11/2022
Event Type  malfunction  
Event Description
It was reported that two sequoia closure tops stripped when the surgeon torqued them.Another closure top was used to complete the surgery and there was no reported patient harm; however, there was a 30 minute delay to the surgery.This is report two of two for this event.
 
Manufacturer Narrative
Without a product return, no product evaluation is able to be conducted.Current information is insufficient to permit a valid conclusion about the cause of this event.If additional information is obtained that adds value to the relevant content of this report and/or a conclusion can be drawn, a follow-up report will be sent.Reference report 3012447612-2022-00250.
 
Manufacturer Narrative
Corrections in d4: lot & udi numbers, d9, h3, and h4.If additional information is obtained that adds value to the relevant content of this report, a follow-up report will be sent.
 
Event Description
It was reported that two sequoia closure tops stripped when the surgeon torqued them.Another closure top was used to complete the surgery and there was no reported patient harm; however, there was a 30 minute delay to the surgery.This is report two of two for this event.
 
Manufacturer Narrative
Device evaluation: visual inspection revealed no damage to the threads or the hex drives.Root cause: root cause was unable to be determined as no product issue was identified.Dhr review: per dhr review, the part was likely conforming when it left zimvie control.Device usage: this device is used for treatment.If additional information is obtained that adds value to the relevant content of this report, a follow-up report will be sent.
 
Event Description
It was reported that two sequoia closure tops stripped when the surgeon torqued them.Another closure top was used to complete the surgery and there was no reported patient harm; however, there was a 30 minute delay to the surgery.This is report two of two for this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPEN IMPLANT CLOSURE TOP, 5.5MM, TI
Type of Device
SEQUOIA PEDICLE SCREW SYSTEM INCLUDING SPEEDLINK II TRANSVERSE CONNECTOR SYSTEM
Manufacturer (Section D)
ZIMMER BIOMET SPINE INC.
10225 westmoor dr.
westminster CO 80021
Manufacturer (Section G)
ZIMMER BIOMET SPINE INC.
10225 westmoor dr.
na
westminster CO 80021
Manufacturer Contact
kim martinez
10225 westmoor dr.
na
westminster, CO 80021
3035144809
MDR Report Key15737079
MDR Text Key305788679
Report Number3012447612-2022-00251
Device Sequence Number1
Product Code NKB
Combination Product (y/n)N
Reporter Country CodeTW
PMA/PMN Number
K131980
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 02/16/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number3301-1
Device Lot NumberABF
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/11/2022
Initial Date FDA Received11/04/2022
Supplement Dates Manufacturer Received12/08/2022
02/03/2023
Supplement Dates FDA Received12/21/2022
02/16/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/08/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient SexFemale
-
-