Model Number 26605 |
Device Problems
Break (1069); Difficult to Advance (2920)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/26/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that hypotube was fractured.A 10.0-31 carotid wallstent stent and a 190cm filterwire ez catheter were selected for use.During the preparation of the filter, it was noted that the hypotube fractured not allowing the delivery of the filter.The stent did not occur over the delivery wire.The procedure was completed with another of the same device.No patient complications were reported.
|
|
Event Description
|
It was reported that hypotube was fractured.A 10.0-31 carotid wallstent stent and a 190cm filterwire ez catheter were selected for use.During the preparation of the filter, it was noted that the hypotube fractured not allowing the delivery of the filter.The stent did not occur over the delivery wire.The procedure was completed with another of the same device.No patient complications were reported.
|
|
Manufacturer Narrative
|
Device evaluated by mfr.: the device was returned for analysis.This carotid device is recommended for use with a 0.014 (0.36mm) guidewire as per carotid wallstent ifu.During the product analysis a boston scientific 0.014 filterwire was successfully inserted through this device with no resistance experienced.The guidewire used by the customer was not returned for analysis.A visual and tactile examination identified no kinks or damage to the delivery system.The device was returned with the stent in the correct position on the device.No issues were noted with the stent.
|
|
Search Alerts/Recalls
|