It was reported that the patient had a reaction to the comfort hard soft splint that was issued.It is unclear when the patient received the device, when the patient first used the device, or when the reaction occurred or resolved.However, it is noted that the patient experienced bumps on upper lip.Also noted was irritation and itching.The device was worn for 10 minutes with the symptoms resolving within the hour.
|
Capa 2023-006.The device was returned, the investigation has been completed and the results are as follows: dhr results: the dhr was reviewed and there was no evidence discovered to indicate that a product defect or non-conformity contributed to the issue.Supplier (erkodent) reviewed the associated material lot and confirmed no material defects or changes could be detected.An allergic reaction is possible, no manufacturing deviation or abnormalities are known.E-pro 4.0 - lot#11847205 (erkoloc-pro) was manufactured from april 04, 2022 and was assigned an expiration date of april 2025.Stock product reviewed results no stock product was available for review since the device was fabricated per physician's prescription only.Investigation methods/results complaint investigator reviewed the returned device.A device was returned in the original case.Roughness - the flange was smooth; internal/external surfaces were smooth.Crack - no major crack was found.Delamination - layers were intact and did not separate.Discoloration - the device was clear with yellow tint and transparent.General cleanliness - the returned device appeared to be clean, very little debris or foreign particles.The returned device was visually inspected and no defect or abnormality was observed.There was no evidence found to indicate that the reported issue was caused by the device itself.Root cause: a root cause for this complaint cannot be explicitly determined.Ifu 9091 rev 4.0 (comfort h/s bite splint instruction for use) states "brush and floss your teeth before use.Rinse mouth well with clean water before inserting the device.If patient uses mouthwash, all traces of mouthwash should be removed by thoroughly rinsing out mouth with water.Rinse bite splint well with clean, cool water before and after use.Clean bite splint with clean, cool water only and let air dry." ifu provides warning "do not clean or soak in mouthwash; do not use denture cleanser, hot water, alcohol, hydrogen peroxide; do not place in direct sunlight".Glidewell research team and namsa conducted a series of testing on a similar thermoformed sleep device (haley) following iso 10993 (biological evaluation of medical devices) and the device was evaluated for potential cytotoxicity, skin irritation, delayed dermal contact sensitization and oral mucosal irritation.The haley test article was thermoformed with layers of erkodent material (erkoloc-pro and erkodur).The test results were listed below and summarized in biocompatibility report for haley sleep device (rpt 9733 rev 1.0) for cytotoxicity testing, the test article extract showed no evidence of causing cell lysis or toxicity.For skin irritation, there was no erythema and no edema observed on the skin of the animals treated with the test article.For sensitization testing, the test article extracts showed no evidence of causing delayed dermal contact sensitization.The test article showed nonirritant to the oral mucosa as compared to the control article.The device materials have been found to be biocompatible through the testing.There was no cytotoxic, sensitization, skin irritation, or oral mucosal irritation found in any of the test articles.
|