MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION EASYTRAK 2 IS-1; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION EASYTRAK 2 IS-1; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D Back to Search Results
Model Number 4542
Device Problems Failure to Capture (1081); High impedance (1291); Connection Problem (2900)
Patient Problem Device Overstimulation of Tissue (1991)
Event Date 01/26/2021
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) and left ventricular (lv) lead showed loss of capture (loc) on the presenting electrogram (egm).In addition, there were multiple high out-of-range pace impedance spikes greater than 3,000 ohms, and it was suspected that the lead-to-head connection was the cause.Technical services (ts) recommended evaluating pace vectors that did not use the ring.Additional information received reported that the patient experienced phrenic nerve stimulation in every other configuration, which is why this vector had been chosen.Thresholds taken that day were 1.5v and output was set to 2v; the safety margin was later increased and trend was programmed on.A twelve lead echocardiogram (ecg) was taken, and capture was confirmed.This crt-d system remains in service and will continue to be monitored.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EASYTRAK 2 IS-1
Type of Device
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
CPI - DEL CARIBE
guidant puerto rico b. v.
#12, rd. #698
dorado PR 00646
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key15739483
MDR Text Key306187671
Report Number2124215-2022-45484
Device Sequence Number1
Product Code NIK
UDI-Device Identifier00802526410901
UDI-Public00802526410901
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
P010012/S024
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial
Report Date 11/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date08/04/2007
Device Model Number4542
Device Catalogue Number4542
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/18/2022
Initial Date FDA Received11/04/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/04/2005
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Age89 YR
Patient SexFemale
-
-