H4 manufacturing date ¿ added.H3 device evaluated by mfg ¿updated.H3 summary attached - updated.D4 expiration date - added.D9 product available to stryker ¿ updated.D9 returned to manufacturer on ¿updated.Due to the automated mes (manufacturing execution system) system there are controls in the manufacturing process to ensure the product met specifications upon release.During visual inspection, the sdw (stent delivery wire) was returned within the introducer sheath, the tip of the introducer sheath was damaged, the distal tip of the sdw had been broken/fractured from the proximal ro marker band and not returned, the sdw was kinked, and the stent was not returned.A functional test was unable to perform due to condition of returned device.The reported event is covered in the device directions for use (dfu).As well, the risk of the reported event is documented in the risk documentation and there are current controls to mitigate the risk of the as reported event.The reported events can be confirmed based on the results of the device analysis.The device did not meet specifications when received for complaint investigation based on the analyzed anomalies noted to the device.It was reported that, flushed and delivered stent through sl-10 and when it reached middle of microcatheter, the stent could not be advanced any more even after several tries.Withdrew the stent system out of microcatheter and then the stent deployed in proximal of sl-10.Additional information received from the customer indicated that the device was prepared for use as per the directions for use, the device was confirmed to be in good condition prior to use on the patient and continuous flush was set up and maintained throughout the clinical procedure.The patient¿s anatomy was described as 'severely tortuous'.The device was returned, and the tip of the introducer sheath was damaged, the sdw was kinked and broken fractured, the damage noted is indicative of the reported event.It is probable that th4e device was damaged during use causing the reported difficulty to advance the stent and the subsequent premature stent deployment.An assignable cause of procedural factors will be assigned to the reported ¿stent difficult/unable to advance or pullback through catheter¿ and ¿stent deployed prematurely during use¿ and to the analyzed ¿stent introducer sheath distal tip damaged¿, ¿sdw broken/fractured during use¿ and ¿sdw kinked/bent¿, as this complaint appears to be associated with a product that met stryker design and manufacturing specifications and was used in accordance with the dfu, but performance was limited due to procedural and/or anatomical factors during use.
|