MEDTRONIC SOFAMOR DANEK USA, INC INFINITY¿ OCCIPITOCERVICAL UPPER THORACIC SYSTEM; APPLIANCE, FIXATION, SPINAL INTERLAMINAL Back to Search Results
Model Number 3600600
Device Problems Device Dislodged or Dislocated (2923); Migration (4003)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/12/2022
Event Type  Injury  
Event Description
Information was received from healthcare provider (hcp) via a manufacturer representative regarding a patient having spinal therapy for unknown indication.It was reported that the infinity set screw popped off the oc plate tulip, rod came out.The product explanted and will be returned. there were no patient symptoms or complications reported as a result of this event.Additional information received from manufacturer representative that the procedure involved was occiput to c3.Patient wanted plate off.Set screw was loose.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
H6: product analysis part# 3600600 lot# rm22c005 visual and optical examination identified that the plate is bent, twisted and damaged.Visual analysis reveals that there is some material deformation on the plate around the screw holes.The material deformation appears to be the result of the screw being screwed in too tight.This is consistent with overload.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFINITY¿ OCCIPITOCERVICAL UPPER THORACIC SYSTEM
Type of Device
APPLIANCE, FIXATION, SPINAL INTERLAMINAL
Manufacturer (Section D)
MEDTRONIC SOFAMOR DANEK USA, INC
4340 swinea rd
memphis TN 38118
Manufacturer (Section G)
MEDTRONIC SOFAMOR DANEK USA, INC
4340 swinea rd
memphis TN 38118
Manufacturer Contact
glen belmer
1800 pyramid place
memphis, TN 38132
6122713209
MDR Report Key15742311
MDR Text Key303148654
Report Number1030489-2022-01014
Device Sequence Number1
Product Code KWP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K163375
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3600600
Device Catalogue Number3600600
Device Lot NumberTMC22C005
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/12/2022
Initial Date FDA Received11/07/2022
Supplement Dates Manufacturer Received01/19/2023
Supplement Dates FDA Received02/14/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age55 YR
Patient SexMale
Patient Weight108 KG
-
-