MEDTRONIC SOFAMOR DANEK USA, INC INFINITY¿ OCCIPITOCERVICAL UPPER THORACIC SYSTEM; APPLIANCE, FIXATION, SPINAL INTERLAMINAL
|
Back to Search Results |
|
Model Number 3600600 |
Device Problems
Device Dislodged or Dislocated (2923); Migration (4003)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/12/2022 |
Event Type
Injury
|
Event Description
|
Information was received from healthcare provider (hcp) via a manufacturer representative regarding a patient having spinal therapy for unknown indication.It was reported that the infinity set screw popped off the oc plate tulip, rod came out.The product explanted and will be returned. there were no patient symptoms or complications reported as a result of this event.Additional information received from manufacturer representative that the procedure involved was occiput to c3.Patient wanted plate off.Set screw was loose.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
H6: product analysis part# 3600600 lot# rm22c005 visual and optical examination identified that the plate is bent, twisted and damaged.Visual analysis reveals that there is some material deformation on the plate around the screw holes.The material deformation appears to be the result of the screw being screwed in too tight.This is consistent with overload.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|